INTENDED USE
The LiveSureTM hCG Rapid Test is intended for detecting the presence of hCG in urine specimens
in a qualitative format sensitive to 20 mIU/ml. This test is for in vitro screening use in
obtaining a visual qualitative result for the early detection of pregnancy.
|
INTRODUCTION
The human chronic gonadotropin (hCG) is a glycoprotein hormone secreted by the developing
placenta shortly after fertilization. The hCG soon appears after conception and only during
early stages of pregnancy. The LiveSureTM hCG Rapid Test is a qualitative, two site sandwich
immunoassay for the determination of hCG antigen. The assay employs a combination of hCG
monoclonal antibodies to selectively detect elevated levels of hCG in urine specimens. The
immunological specificity of the test kit virtually eliminates cross reactivity in interference
from the structurally related glycoprotein hormones hFSH, hLH and hTSH at physiological levels.
|
PRINCIPLE
The LiveSureTM hCG Rapid Test is a chromatographic immunoassay (CIA) for the rapid qualitative
determination of hCG in specimens. The membrane was precoated with hCG specific antibodies on
the test region. During the test, the specimen is allowed to react with the hCG monoclonal
antibody-colloid gold conjugate which was pre-dried on the test strip. The mixture then moves
upward on the membrane chromatographically by the capillary action. For a positive specimen,
the conjugate binds to the hCG forming an antibody-antigen complex. This complex binds to the
hCG antibody as a captured reagent on the test region and produces a colored band when hCG
concentration is equal to or greater than 20 mIU/ml. Absence of this colored band in the test
region suggests a negative result. To serve as a procedural control, a colored band at control
region will always appear regardless of the presence of hCG.
|
STORAGE AND STABILITY
The LiveSureTM hCG Rapid Test can be store refrigerated or at room temperature (2-28ºC) in tightly
closed container. Avoid freezing.
|
PRECAUTIONS
- For in vitro diagnostic use.
- Do not use after expiration date.
- Test strip should remain in securely closed container until ready for use.
- The reagents in this kit contain sodium azide which may react with lead or copper plumbing
to form potentially explosive metal azide build up. Specimens should be considered hazardous
and handled appropriately.
|
SPECIMEN COLLECTION
Collect and store specimens following standard clinical procedures.
- The urine specimen must be collected in a clean dry container, either plastic or glass,
without preservative. No centrifugation or filtration of urine is required.
- Specimens collected at any time may be used. However, the first morning urine generally
contains the highest concentration of hormone.
- If specimens cannot be tested after collection, they should be stored refrigerated at 2-8°C.
|
ASSAY PROCEDURE
For Dipstick Test
- Remove the strip from container and label the strip with specimen identification.
- Carefully place the white end (sample pad) of the reaction strip into the specimen vial. Make sure that the absorbent patch white end of the strip) is underneath the surface of the sample liquid. A 10 to 30 second dip into the sample is sufficient.
- Remove end of the test strip from the reaction vial and start the watch or timer.
- Within 5 minutes, a colored band will appear at the top of the test region, indicating the reaction is completed.
- Observe the result within 5 minutes, no longer than 10 minutes.
|
INTERPRETATION OF RESULT
Negative: Only one colored band appears on the control region (C). No colored band in the test region (T).
Positive: In addition to the control band, a distinct colored band also appears in the test region (T).
Invalid: If no bands appear after 10 minute, the result is invalid. The protocol may not have been performed correctly, or the test is deteriorated. The assay should be repeated using a new test device.
Note: Do not interpret result after 10 minutes.
|
LIMITATIONS
- If a specimen is too diluted (i.e. low specific gravity), it may not contain representative levels of hCG.
If pregnancy is still suspected, a first morning urine should be obtained from the patient and tested. HCG
concentrations less than 20 mIU/ml will be detected as negative.
- A number of conditions other than pregnancy including trophoblastic disease, proteinuria, hematuria,
choriocarcinoma, ovarian and testicular teratomas cause elevated levels of hCG. These diagnoses should be
considered if appropriate to the clinical evidence.
- As with all diagnostic tests, a definitive clinical diagnosis should not be based on the results of a
single test, but should only be made by the physician after all clinical and laboratory findings have been
evaluated.
|
PERFORMANCE AND CHARACTERISTIC
Sensitivity:
The analytical sensitivity of the LiveSure™ hCG Rapid Test has been set at 20 mIU/ml. The 40 mIU/ml Positive Control (calibrated to the 2nd International Standard) was designed as the cutoff for the test because hCG concentrations in this range are usually achieved the 2nd week post conception. The test will yield a positive result on the first day of missed menstrual period.
Specificity:
The specificity was determined from cross reaction studies with known amounts of Luteinizing Hormone (hLH), Follicle Stimulating Hormone (hFSH), and Thyroid Stimulating Hormone (hTSH). 300 mIU/ml hLH, 1000mIU/ml hFSH and 1000 mIU/ml hTSH all gave negative results.
Accuracy:
A study was performed using a total of 251 randomly selected urine specimens. These specimens were assayed with hCG Rapid Test and a similar commercially available qualitative visual pregnancy test. The results indicated a complete agreement (99 positive specimens and 152 negative specimens).
Interference Testing:
The following substances at certain concentrations do not interfere with the hCG Rapid Test in the assay.
| Acetaminophen | 20 mg/dl | Acetylsalicylic Acid | 20 mg/dl |
| Ascorbic Acid | 20 mg/dl | Atropine | 20 mg/dl |
| Caffeine | 20 mg/dl | Gentesic Acid | 20 mg/dl |
| Glucose | 2.0 g/dl | Hemoglobin | 1.0 mg/dl |
|
REFERENCES
- Pan Probe Biotech, Inc., package insert.
- Cart, K. J.: J. Clin. Endocrinol. Metab., 1975, 40:537
- Brauntein, G.D.: Am. J. Obstet. Gynecol., 1976, 126:678
- Batzer, F.R.: Fertility & Sterility, 1980, 34:1
- Engvall, E.: Method in Enzymology, 1980, 70:419
- Wilcox,A.J.: N. England J. Med. 1988, 319:189
- Dawood, M.Y.: Ob. Gyn., 1990, 126:678
|
|
 |