INTENDED USE
Pan Probe LiveSure™ Multi-Drug 5 Panel Test is a diagnostic qualitative lateral flow immuno-chromatographic
urinary assay for the rapid detection of drug of abuse in human urine at cut-off levels specified by SAMHSA in a
convenient one step multiple test strip format. Test strips included in the LiveSure™ Multi-Drug 5 Panel Test are
Opiate, Cocaine, Amphetamine, Marijuana (THC) and Methamphetamine. This test is designed to obtain a visual,
qualitative result and intended for professional use only. It is not intended for quantitative results, or for Over
the Counter (OTC) sale to lay persons.
The test provides only preliminary analytical data. A more specific alternative method must be used in
order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) has been established as
the preferred confirmatory method by the Substance Abuse and Mental Health Services Administration (SAMHSA).
Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly
when preliminary positive results are indicated.
|
SUMMARY AND EXPLANATION OF TEST
The LiveSure™ Multi-Drug 5 Panel Test is a convenient holder which permits 5 individual determinations in
one dip using the LiveSure™ drugs of abuse test strips. It is easy, fast and visually indicates competitive binding
immunoassay results for screening without the need for instrumentation to arrive at a determination. There is no
intercommunication between strips and each test strip functions completely independently. Each individual test employs
unique specific antibodies to selectively identify the individual drug of abuse in test samples.
|
PRINCIPLE OF THE PROCEDURE
The LiveSure™ Multi-Drug 5 Panel Test contains sample a reaction unit, a colloidal gold conjugate color unit
pre-labeled with specific antibody, and a chromatographic membrane unit pre-coated with drug conjugate in the test region.
When the free drug is absent or less than the detection cut-off level in the urine, the colored antibody-colloidal gold
conjugate is free to bind to the immobilized drug conjugate pre-coated on the chromatographic membrane test region. A pink
band will form in the test region indicating a negative result. When the drug is present in the urine at or exceeding the
detection cut-off level, it will bind to the limited epitomes on the colored antibody-colloidal gold conjugate, which
completely competes with drug conjugate in the test region. As a result, it appears as an absence of the color band in
the test region indicating a positive result. A control band is designed to indicate that the test has been performed
properly. It should always show up in the control region regardless of the drug and metabolite presence, demonstrating
that the assay system is functioning correctly. Summarizing, negative urine will produce two colored bands, one in both
the control region and test region, while positive urine produces only one band in the control region.
|
REAGENTS AND MATERIALS SUPPLIED
- LiveSure™ Multi-Drug 5 Panel Test device, containing sample reaction unit, colloidal gold conjugate color unit, and chromatographic membrane unit
- Test Instructions
|
MATERIALS REQUIRED BUT NOT PROVIDED
- Urine specimen collection container
- Timer or clock
|
PRECAUTIONS
- For in vitro diagnostic use.
- Avoid cross contamination of urine samples by using a new specimen collection container for each urine sample.
- Urine specimens may be potentially infectious. Avoid contact with skin by wearing gloves and proper laboratory attire. Properly handle and dispose of all used test devices in an appropriate manner.
- Do not use a test device beyond the expiration date imprinted on the outside of the foil pouch.
|
STORAGE AND STABILITY
The test devices supplied can be stored at room temperature or refrigerated (2-28°C) and will be stable until the expiration date. Do not freeze.
|
SAMPLE COLLECTION AND PREPARATION
The fresh urine samples should be collected in a clean dry container such that testing may be performed
during the same day. Urine specimen may be refrigerated at 2-8°C for 48 hours, or frozen at -20°C for longer
period of time prior to assaying. Specimens that have been refrigerated or frozen must be equilibrated to
room temperature before testing. Urine samples exhibiting visible precipitates should be filtered, centrifuged
or allowed to settle so that clear aliquots can be obtained for testing.
|
ASSAY PROCEDURE
- Do not open foil pouch until ready to begin testing. Refrigerated test device should be allowed to come
to room temperature (15-28°C) before opening the pouch.
- Remove the test device from the sealed foil pouch by tearing along the notch.
- Remove the protective cap and immerse the strips in urine with the arrow end pointing towards the urine
sample for at least 10 seconds. Do not introduce urine above the maximum level line as indicated by the arrows.
- Replace the protective cap over the exposed strips and lay the test device on a clean flat surface.
Read the test result at 5 minutes.
IMPORTANT: In order to prevent an incorrect reading, do not read the test results after more than 10
minutes. If the test is read after 10 minutes, the intensity of the colored lines may change or a new line may
appear. To avoid confusion, discard the test device after interpreting the result.
|
INTERPRETATION OF RESULTS
Negative: Two colored bands appear, one in the control region and one in the test region. A
negative result indicates that the free drug is absent or less than the detection level of the test.
Positive: Only one colored band appears in the control region with no apparent band in the test
region. A positive result indicates that the free drug is present in the urine at or exceeding the detection
level of the test.
Invalid: No band appears in the control region, or a band appears in the test region only. An
invalid result may be due to improper assay procedures or damage of the device. The assay is inconclusive
and the specimen should be repeated using a new test device.
Note: The test band intensity may be weaker or stronger than that of the control band.
TEST DEVICE
|
 |
 |
 |
| Positive(+) |
|
Negitive(-) |
|
QUALITY CONTROL
Good laboratory practice recommends the use of control materials to ensure proper test
device performance and reliability. When testing the quality control standards, use the same assay
procedure as with a urine sample. The SAMHSA recommended guidelines for drug of abuse screening
test device indicate that controls should contain the drug of abuse analyte at levels at least
25% above the SAMHSA cut-off values.
|
AFTER TESTING
Urine specimens may be potentially infectious. Avoid contact with skin by wearing gloves
and proper laboratory attire. Properly handle and dispose of all tested devices in an appropriate manner.
The residual urine should be disposed of in a medically approve manner after the completion of all
testing including the confirmatory assay.
|
PERFORMANCE CHARACTERISTICS
SENSITIVITY:
Pan Probe LiveSure™ Multi-Drug 5 Panel Test has been designed for the detection of drugs of
abuse and their metabolites in urine at the detection sensitivity of levels as suggested for
immunoassay methods by SAMHSA.
| Amphetamine | 1,000 ng/ml |
| Methamphetamine | 1,000 ng/ml |
| Cocaine | 300 ng/ml | | Marijuana(THC) | 50 ng/ml |
| Opiate | 2,000 ng/ml |
PRECISION:
The results of within-lot reproducibility analytical studies clearly showed excellent repeatability for
all 3 batches of positive and negative urine samples, using one lot of the LiveSure™ Multi-Drug 5 Panel Test.
The results of these tests clearly demonstrate that there is no appreciable inter-lot variation when testing
both positive and negative spiked samples across three (3) different lots of LiveSure™ Multi-Drug 5 Panel Test.
SPECIFICITY:
The specificity of the devices was tested by compounds related to and also not associated with drugs of abuse
prepared in drug-free normal human urine. The following compounds produced positive results in the test device
at or greater than the concentrations listed below:
| Opiates | Morphine | 2,000 ng/ml |
| | Codeine | 2,000 ng/ml |
| | Ethyl morphine | 2,000 ng/ml |
| Cocaine | Benzoylecgonine | 300 ng/ml |
| | Cocaine | 300 ng/ml |
| Amphetamine | D-(+)-Amphetamine | 1,000 ng/ml |
| | (±) 3,4-Methylenedioxyamphetamine | 5,000 ng/ml |
| | L-Amphetamine | 10,000 ng/ml |
| Marijuana(THC) | 11-nor-D9-THC-9-COOH | 50 ng/ml |
| | 11-nor-D8-THC-9-COOH | 250 ng/ml |
| | Cannabinol | 50,000 ng/ml |
| | D8-THC | 25,000 ng/ml |
| | D9-THC | 15,000 ng/ml |
| | 11-hydroxy-D9-THC | 10,000 ng/ml |
| Methamphetamine | (+)-Methamphetamine | 1,000 ng/ml |
| | (±) 3,4-Methylenedioxymethamphetamine | 2,500 ng/ml |
| | Procaine | 10,000 ng/ml |
| | (-)-Methamphetamine | 25,000 ng/ml |
| | D-(+)-Amphetamine | 50,000 ng/ml |
| | Chloroquine | 50,000 ng/ml |
| | (±)-Ephedrine | 50,000 ng/ml |
| | *-Phenylethylamine | 50,000 ng/ml |
| | Ranitidine | 50,000 ng/ml |
The following compounds were found not to have any impact on the negative results when tested at concentrations up to 100,000 ng/ml.
Acetaminophen
Acetone
Albumin
Amikacin
Amitriptyline
Ampicillin
Arterenol
Aspartame
Aspirin (Acetylsalicylic Acid)
Benzoic Acid
Bilirubin
Caffeine
Chloroquine
(+)-Chlorpheniramine
(+/-)-Chlorpheniramine
Cimetidine
Deoxyephedrine
Dexbrompheniramine
Dextromethorphan
Diphenylhydantoin
Doxylamine
Erythromycin
Ethanol
|
Furosemide
Glucose
Guaiacol Glyceryl Ether
Hemoglobin
Histamine
Imipramine
Indomathacin
(+/-)-Isoproterenol
Methyphenidate
(+/-)-Norephedrine
Oxalic Acid
Pendimetrazine
Penicillin-G
Propanol
Phentermine
Quinine
Quinidine
Raboflavin
Ranitidine
Sodium Chloride
Sulindac
Tyramine
Vitamin C
|
|
ASSAY COMPARISONS & EQUIVALENCY
Each of the test strips eligible for inclusion into the holder have been subjected to strict
testing prior to FDA clearance of the individual test strips. In independent clinical trials each analyte
strip was subjected to evaluation involving a comparison between the strip, a predicate device (EMIT®II)
and GC/MS.
Accuracy and equivalency comparisons of each LiveSure™ Multi-Drug 5 Panel Test was first tested
in individual urine samples in-house and subsequently in a clinical trial of individual urines submitted to
a SAMHSA certified laboratory. All positive samples by either screening method were confirmed by GC/MS. The
results of that testing have been tabulated below for all five analytes.
| Opiates |
| LiveSure™ | EMIT® II (+) | EMIT® II ( - ) | Row Totals |
| (+) | 146 | 0 | 146 |
| ( - ) | 6 | 331 | 337 |
| Col. Totals | 152 | 331 | 483 |
Then compared to Emit® II Assay, the percent agreement with LiveSure™ Morphine Screen Test strip positive
samples and 146/152 or 96.1%. Negative samples recovered at 331/331 or 100%, while the overall relative accuracy
obtained was 477/483 or 98.8%.
| LiveSure? | GC/MS (+) | GC/MS ( - ) | Row Totals |
| (+) | 100 | 0 | 100 |
| ( - ) | 0 | 281 | 281 |
| Col. Totals | 100 | 281 | 381 |
Then compared to the GC/MS data, the relative sensitivity or percent agreement with LiveSure™ Morphine
Screen Test Strip positive samples with the external clinical study was 100/100 or 100%. Negative samples recovered
a relative specificity of agreement of 281/281 or 100%. Finally, the overall relative accuracy obtained was 381/381 or 100%.
| Cocaine |
| LiveSure? | EMIT® II (+) | EMIT® II ( - ) | Row Totals |
| (+) | 162 | 0 | 162 |
| ( - ) | 5 | 322 | 327 |
| Col. Totals | 167 | 322 | 489 |
When compared to Emit® II Assay, the percent agreement with LiveSure™ Cocaine Screen Test Strip positive
samples and 162/167 or 97.0%. Negative samples recovered at 322/322 or 100%, while the overall relative accuracy
obtained was 484/489 or 99.0%.
| LiveSure? | GC/MS (+) | GC/MS ( - ) | Row Totals |
| (+) | 101 | 4 | 105 |
| ( - ) | 0 | 272 | 272 |
| Col. Totals | 101 | 276 | 377 |
When compared to the GC/MS data, the relative sensitivity or percent agreement with LiveSure™ Cocaine Screen
Test Strip positive samples with the external clinical study was 101/101 or 100%. Negative samples recovered a relative
specificity of agreement of 272/276 or 98.6%. Finally, the overall relative accuracy obtained was 373/377 or 98.9%
| Amphetamine |
| LiveSure? | EMIT® II (+) | EMIT® II ( - ) | Row Totals |
| (+) | 141 | 0 | 141 |
| ( - ) | 3 | 220 | 223 |
| Col. Totals | 144 | 220 | 364 |
When compared to Emit® II Assay, the percent agreement with LiveSure™ Amphetamine Screen Test strip positive
samples and 141/144 or 97.9%. Negative samples recovered at 220/220 or 100%, while the overall relative accuracy obtained was 361/364 or 99.2%.
| LiveSure? | GC/MS (+) | GC/MS ( - ) | Row Totals |
| (+) | 101 | 4 | 105 |
| ( - ) | 0 | 272 | 272 |
| Col. Totals | 101 | 276 | 377 |
When compared to the GC/MS data, the relative sensitivity or percent agreement with LiveSure™ Amphetamine
Screen Test Strip positive samples with the external clinical study was 94/94 or 100%. Negative samples recovered a relative
specificity of agreement of 161/163 or 98.8%. Finally, the overall relative accuracy obtained was 255/257 or 99.2%.
| Marijuana (THC) |
| LiveSure? | EMIT® II (+) | EMIT® II ( - ) | Row Totals |
| (+) | 160 | 0 | 160 |
| ( - ) | 8 | 341 | 349 |
| Col. Totals | 168 | 341 | 509 |
When compared to Emit® II Assay, the percent agreement with LiveSure™ THC Screen Test Strip positive
samples and 160/148 or 95.2%. Negative samples recovered at 341/341 or 100%, while the overall relative accuracy
obtained was 501/509 or 98.4%.
| LiveSure? | GC/MS (+) | GC/MS ( - ) | Row Totals |
| (+) | 107 | 5 | 112 |
| ( - ) | 0 | 283 | 283 |
| Col. Totals | 107 | 288 | 395 |
When compared to the GC/MS data, the relative sensitivity or percent agreement with LiveSure™ THC
Screen Test Strip positive samples with the external clinical study was 107/107 or 100%. Negative samples recovered
a relative specificity of agreement of 283/288 or 98.3%. Finally, the overall relative accuracy obtained was
390/395 or 98.7%.
| Methamphetamine |
| LiveSure? | EMIT® II (+) | EMIT® II ( - ) | Row Totals |
| (+) | 205 | 3 | 208 |
| ( - ) | 1 | 211 | 212 |
| Col. Totals | 206 | 214 | 420 |
When compared to Emit® II Assay, the percent agreement with LiveSure™ Methamphetamine Screen Test
Strip positive samples and 205/206 or 99.5%. Negative samples recovered at 211/214 or 98.6%, while the overall
relative accuracy obtained was 416/420 or 99.1%.
| LiveSure? | GC/MS (+) | GC/MS ( - ) | Row Totals |
| (+) | 133 | 5 | 138 |
| ( - ) | 0 | 162 | 162 |
| Col. Totals | 133 | 167 | 300 |
When compared to the GC/MS data, the relative sensitivity or percent agreement with LiveSure™
Methamphetamine Screen Test Strip positive samples with the external clinical study was 133/133 or 100%.
Negative samples recovered a relative specificity of agreement of 162/167 or 97.0%. Finally, the overall
relative accuracy obtained was 295/300 or 98.3%.
|
LIMITATIONS OF PROCEDURE
- This diagnostic test is designed to be used for the medical or forensic in vitro detection of drugs of abuse in
urine only.
- This diagnostic test provides a qualitative screening for drugs of abuse in urine only, and is not to be used for
quantitative determination of drug concentration in urine. A positive result with this test indicates only the presence
of a drug or metabolite, does not indicate intoxication, and should be confirmed by GC/MS.
- There is a possibility of false positive results due to the assay procedural errors or other substances present
in the urine as interfering factors, some of which are listed.
- See the specificity lists of substances that will produce positive results at high levels, versus those
substances that do not interfere with test performance.
- Adulterants, such as bleach or other strong oxidizing agents, when added to urine specimens, may produce erroneous
test results regardless of the analysis method. If an adulteration is suspected, another urine specimen should be used.
|
BIBLIOGRAPHY
- Pan Probe Biotech, Inc., package insert.
- Baselt, R. C. Disposition of Toxic Drugs and Chemicals in Man, Biomedical Publications, Davis, CA, 1982.
- Urine testing for Drugs of Abuse. National Institute on Drug Abuse (NIDA), Research Monograph 73, 1986.
- Fed. Register, Department of Health and Human Services, Mandatory Guidelines for Federal Workplace Drug
Testing Programs, 53, 69, 11970-11979, 1988.
- McBay, A. J. Clin. Chem. 33, 33B-40B, 1987.
- Gilman, A. G., and Goodman, L. S. The Pharmacological Basis of Therapeutics, eds. MacMillan Publishing,
New York, NY, 1980.
|
|
 |